Beurer medical BM 44 Manual De Instrucciones página 96

Tensiómetr
Ocultar thumbs Ver también para BM 44:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 36
Table 4
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and the BM 44
The BM 44 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of
the BM 44 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the BM 44 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be deter-
mined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from struc-
tures, objects and people.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1,2
0,12
0,38
1,2
3,8
12
96
(m)
80 MHz to 800 MHz
d = 1,2
0,12
0,38
1,2
3,8
12
800 MHz to 2,5 GHz
d = 2,3
0,23
0,73
2,3
7,3
23

Publicidad

Tabla de contenido
loading

Tabla de contenido