Setup Menu Table - Thermo Scientific Orion Star Manual Del Usario

Ocultar thumbs Ver también para Orion Star:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Meter Setup

Setup Menu Table

The following table is for the complete line of Orion Star and Star Plus meters .
Meters may not include all of the options listed in this table .
Top Line
Middle Line
PH
rES
PH
bUF
ISE
rES
ISE
UnIt
ISE
rAng
ISE
nLIn
C0nd
tC
C0nd
C0EF
C0nd
tdSF
C0nd
CELL
C0nd
trEF
C0nd
tyPE
See the RDO Optical Dissolved Oxygen Setup Menu section for details on the RDO probe information menus.
d0
rES
d0
rES
d0
bAr
d0
PrES
d0
SAL
d0
SALF
d0
CALt
d0
LIFE
dUE
PH
dUE
0rP
dUE
ISE
dUE
C0nd
dUE
d0
EN-12
Bottom Line
0.1, 0.01, 0.001
USA, EUr0
1, 2, 3
m, mgL, PEr, PPb, n0nE
L0w, HIgH
AUt0, 0FF
0FF, LIn, nLF
0.0 to 10.0
0.00 to 10.0
0.001 to 199.0
5, 10, 15, 20, 25
Std, 1, 2, 3, 4, 5, 6, 7, USP
0.1, 1 % sat
0.01, 0.1 mg/L
AUt0, mAn
450.0 to 850.0
AUt0, mAn
0 to 45
AIr, H20, mAn, SEt0
365 to 0
0 to 9999
0 to 9999
0 to 9999
0 to 9999
0 to 9999
Thermo Scientific Orion Star
Setup Menu Description (default setting, method specific)
pH measurement resolution (0.01, yes)
pH buffer set for automatic buffer recognition during calibration,
USA buffers are 1.68, 4.01, 7.00, 10.01, 12.46 and EUrO buffers
are 1.68, 4.01, 6.86, 9.18 (USA, yes)
ISE measurement resolution in significant figures (1, yes)
ISE measurement units (PPb, yes)
ISE concentration range for calibration stability criteria (HIgH, yes)
ISE automatic blank correction for low-level calibration (AUt0, yes)
Conductivity temperature compensation type, LIn is for linear,
nLF is for non-linear pure water samples (LIn, yes)
Conductivity temperature compensation coefficient in % change
in conductivity per °C, appears if LIn was selected for tC (2.1, yes)
Conductivity TDS factor value (0.49, yes)
Conductivity default cell constant value for automatic conductivity
calibration mode (0.475, yes)
Conductivity reference temperature (25, yes)
Conductivity cell type and selectable range. Std is for standard,
USP is for planar, 1 through 7 for fixed ranges (Std, yes)
DO % saturation measurement resolution (0.1, yes)
DO mg/L measurement resolution (0.01, yes)
DO barometric pressure compensation type (AUt0, yes)
DO manual barometric pressure compensation value, appears
if mAn was selected for bAr (760.0, yes)
DO salinity correction type (AUt0, yes)
DO manual salinity correction value, appears if mAn was selected
for SAL or a DO meter without a conductivity mode is used (0, yes)
DO calibration type (AIr, yes)
Optical cap replacement countdown in days – 3, 4 and 5 Star Plus
RDO meters only (Set by RDO probe, no)
pH calibration alarm value in hours, 0 is off (0, yes)
ORP calibration alarm value in hours, 0 is off (0, yes)
ISE calibration alarm value in hours, 0 is off (0, yes)
Conductivity calibration alarm value in hours, 0 is off (0, yes)
DO calibration alarm value in hours, 0 is off (0, yes)
and Star Plus Meter User Guide

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Star plus meter

Tabla de contenido