EN-24
b)
While the unit is running a program, the left
knob (C) and the right knob (D) will be
disabled.
19. To stop a program while it is running, flip the
switch (J) on the side of the unit. Flip the switch
(J) back to on and the unit will return to the main
operating screen and be ready for normal use.
2.11 Using the SmartRate™ Feature
SmartRate™ allows the user to change the rate at which the heater and/or stirrer increases temperature or speed
respectively.
1. Icons
a)
The turtle icon (
temperature or speed.
b)
The rabbit icon (
quicker pace than the default setting.
CAUTION: In this faster mode the heater or stirrer will reach temperatures or speeds greater than the
setting prior to stabilization.
The double dash icon (--) indicates that the SmartRate™ feature is turned off. (This will only be
c)
displayed in the SmartRate™ settings menu.)
2. To control the SmartRate™ Feature, the unit must
first be in the top level of the settings menu.
3. Rotate the left knob (C) to scroll to the "RAMP"
setting.
4. Briefly press the left knob (C) to enter the
SmartRate™ settings menu.
5. To change the SmartRate™ of the heater, briefly
press the left knob (C).
The icon for the current SmartRate™ setting will
begin to blink.
6. Rotate the left knob (C) to scroll to the desired
SmartRate™ setting for the heater.
) allows for a slower, more precise rate and prohibits overshoot of the target
) allows the heater and/or stirrer to reach the target temperature or speed at a
Guardian 7000 Series