SYMBOLS
Symbol
ON/OFF
MAINTENANCE
The device is frangible and must be handled with care.Gel must be wiped
from the probe after use. These precautions will prolong the life of the
unit. The user must check that the equipment does not have visible evi-
dence of damage that may affect patient safety or Pocket Fetal Doppler
capability before use. The recommended inspection interval is once per
month or less. If damage is evident, replacement is recommended before
use. The device should undergo periodic safety testing to insure proper
patient isolation from leakage currents. This should include leakage cur-
rent measurement. The recommended testing interval is once every two
years or as specified in the institution's test and inspection protocol. The
accuracy of FHR is controlled by the equipment and can not be adjusted
by user. If the FHR result is distrustful, please use other methods such as
stethoscope to verify immediately or contact local distributor or manu-
facture to get help.
Description
Type B product
Attention
Headset
Volume switch
Power ON/OFF
Heart Rate
23