8. Technical Specifications

Model
Type
Temperature Measurement Range Ear: 82.4 °F - 109.4 °F (28.0 °C - 43.0 °C)
Measurement Accuracy
Measurement Duration
Operating conditions
Ambient temperature is not within mea-
surement range.
Thermometer not placed in ear cor-
rectly.
Large fluctuation in ambient tempera-
ture.
Thermometer is not working properly.
Thermometer is in Surface Temperature
mode and Protective Cover is cover-
ing sensor.
In-ear measurement:
the measured temperature is over
109.4 °F (43 °C).
Object measurement:
the measured temperature is over
212 °F (100 °C).
In-ear measurement:
the measured temperature is below
82.4 °F (28 °C).
Object measurement:
the measured temperature is below
-4 °F ( -20 °C).
Batteries low or depleted.
AFT 58
FT 58
Surface temperature: -4 °F - 212 °F ( -20 °C - 100 °C)
Ambient temperature: 32 °F - 122 °F (0 °C - 50 °C)
Ear: ±0.4 °F (0.2 °C) from 95.9 °F - 107.6 °F (35.5 °C - 42.0 °C);
±0.5 °F (0.3 °C) below 95.9 °F (35.5 °C) or above 107.6 °F (42.0 °C)
Object temperature: ±4 °F (2 °C) or 4%
Room temperature: ±4 °F (2 °C)
Approx. one second
50 °F - 104 °F (10 °C - 40 °C),
Humidity 15 - 85% (non condensing)
Place the thermometer in an ambient
temperature between 50° F and 104° F
(10° C - 40° C) for at least 30 minutes.
Please observe the notes in the chapter
"Measuring". Perform the measurement
again.
Store the device for at least 30 minutes
in the room where the measurement will
be taken.
Place the ear thermometer in a room with
an ambient temperature between 50° F
and 104°F (10°C and 40°C) for at least
30 minutes.
Replace batteries and retry. If problems
persist, contact customer service.
Press the Protective Cover Removal Lever
to remove cover.
Only use thermometer within measure-
ment range.
Ensure that the device and the protective
cover are fully functioning. Perform the
measurement again.
Only use thermometer within measure-
ment range.
Ensure that the device and the protective
cover are fully functioning. Perform the
measurement again.
Replace batteries.