Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Do not open the device. It uses delicate electronic components and an intricate
air unit that could be damaged. If you can not fix the problem using the
troubleshooting instructions, request service from your supplier or from the A&D
service group. The A&D service group will provide technical information, spare
parts and units to authorized suppliers.
The technical testing procedures, which should be done at least every three
years, can be performed either by the manufacturer or by an authorized repair
service in accordance with the regulations governing manufacturing of medical
products.
Type
Measurement method
Measurement range
Measurement accuracy
Power supply
Arm circumference
Classification
Clinical test
Operating condition
Storage condition
Dimensions
Weight
Note
Specifications are subject to change without prior notice.
Maintenance

Technical Data

UA-702
Oscillometric measurement
Pressure : 20 - 280 mmHg
Pulse :
Pressure : ±3 mmHg or 2%, whichever is greater
Pulse :
4 x 1.5V batteries (R6P, LR6, or AA)
22 - 32 cm
Type BF
According to ANSI / AAMI SP-10 1987
? 10 °C to ? 40 °C / 30%RH to 85 %RH
? 20 °C to ? 70 °C / 30%RH to 85 %RH
Approx. 135 [W] x 45 [H] x 105 [D] mm
Approx. 200 g, excluding batteries
ENGLISH 12
40 - 200 beats / minute
±5%

Publicidad

Tabla de contenido
loading

Tabla de contenido