2.2.4 Resistance Temperature Detecting Device With Protecting Well; (Separative Element) For Zone 1; Thermocouple Temperature Detecting Device With/Without Protecting Well For The Use In Zone 0/Zone 1 - ABB SensyTemp MI Instrucciones De Montaje

Tabla de contenido

Publicidad

2.2.4 Resistance temperature detecting device with protecting well

(separative element) for zone 1

∅ 3mm
Temperature
category/
temperature
[ °C ]
Power Pi ( Po )
[W]
DMW
0
T6/ 80
0,75
T5/ 95
1,02
T4/130
1,50
T3/195
-
T2/290
-
T1/440
-
Tabelle 2-4
DMW: wire multiplier resistor, SMW: layer multiplier resistor
2.2.5 Thermocouple temperature detecting device with/without
protecting well for the use in zone 0/zone 1
∅ 3mm
Temperature
category/
temperature
[ °C ]
Power Pi ( Po )
[W]
Without
protecting
well
0
T6/64
1,22
T5/76
1,50
T6/80
1,50
Tabelle 2-5
GB 8
Surface
temperature
[°C]
SMW
DMW
0
40
0,60
80
0,83
95
1,35
120
1,50
-
-
-
-
-
Surface
temperature
[°C]
With
Without
protecting
protecting
well
well
0
40
1,50
64
-
68
1,50
68
SensyTemp MI, IS
∅ 6 mm
Power Pi ( Po )
[W]
SMW
DMW
40
0
80
1,26
95
1,50
130
-
140
-
-
-
-
-
∅ 6 mm
Power Pi ( Po )
[W]
With
Without
protecting
protecting
well
well
40
0
64
1,50
-
-
64
1,50
Surface
temperature
[°C]
SMW
DMW
0
40
1,16
80
1,50
87
-
-
-
-
-
-
-
-
Surface
temperature
[°C]
With
Without
protecting
protecting
well
well
0
40
1,50
61
-
-
1,50
61
42/10-59 XU
SMW
40
80
92
-
-
-
-
With
protecting
well
40
57
-
57

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Sensytemp is

Tabla de contenido