The following instructions allow the correct use of the printer supplied with the machine.
By connecting the machine to the power supply, the printer will emit a beep and start and the led "
will light up.
For feeding the paper press the key "
•
"ON LINE" mode : as soon as the acquisition time has passed, the machine prints and saves the values
sensed by the probes.
•
Preset Time mode : as soon as the acquisition time has passed, the machine saves the values sensed
by the probes and prints them at the preset time.
Possibile alarms and signalling are displayed :
When the "ON/OFF" led lights up: the machine is on.
When the "ALARM" led flashes, it indicates that the alarm is active.
When the "SET" Led lights up, it indicates that the configuration parameters are being set.
When the "SET" Led flashes, it indicates that the value of a configuration parameter is been modified
When the "TIMER" light up, it indicates that the preset time mode is active.
When "Printing..." is displayed, it indicates that the "ON LINE" printing mode is active.
When "Recording..." is displayed, it indicates that the preset time printing mode is active.
"Memory 90%...99%" indicates that the memory is about running out.
When "AN1 ERR" (probe 1 error) is displayed, it may indicate one of the following errors: probe 1 type
uncorrectly linked, machine-probe 1 uncorrectly linked, the value sensed by the probe 1 is over the
measure field.
When "AN2 ERR" (probe 2 error) is displayed, it may indicate one of the following errors: probe 2 type
uncorrectly linked, machine-probe 2 uncorrectly linked, the value sensed by the probe 2 is over the
measure field.
When "AN1 AH" (probe 1 maximum error alarm) is displayed, it indicates that the value sensed by the
probe 1 is out of the set limit.
When "AN1 AL" (probe 2 minimum error alarm) is displayed, it indicates that the value sensed by the
probe 2 is out of the set limit.
When AN2 AH" (probe 2 maximum error alarm) is displayed, it indicates that the value sensed by the
probe 2 is out of the set limit.
When "AN2 AL" (probe 2 maximum error alarm) is displayed, it indicates that the value sensed by the
probe 2 is out of the set limit.
When "Memory Full" is displayed (running out memory alarm), it indicates that memory is running out .
39C1000
TECHNICAL PRINTER USERS GUIDE
". (1)
SIGNALLINGS AND ALARMS
START PROOF
PAPER FEED
PRINTING MODE
"