Error Messages/Malfunctions - Procheck WW1YB-3 Manual De Instrucciones

Tensiómetro de muñeca
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

☞ Further information
Measurements should not occur soon after each other, since otherwise, the results will be falsi-
fied. Wait therefore for several minutes in a relaxed position, sitting or lying, before you repeat
a measurement.

6. Error messages/malfunctions

If an error occurs during a measurement, the
measurement is discontinued and a corresponding error
code is displayed (Example: Error No. 2).
Error No. Possible cause(s)
ERR 1
No pulse has been detected.
ERR 2
Unnatural pressure impulses influence the measurement result. Reason: The arm
was moved during the measurement (Artefact).
ERR 3
The inflation of the cuff takes too long. The cuff is not correctly seated.
ERR 5
The measured results have indicated an unacceptable difference between the sys-
tolic and diastolic pressure. Take another measurement, carefully following the
instructions. Contact your doctor, if you still obtain unusual results.
HI
The pressure in the cuff is too high (over 300 mmHg) OR the pulse is too high (over
200 beats per minute). Relax for 5 minutes and repeat the measurement.*
LO
The pulse is too low (less than 40 beats per minute). Repeat the measurement.*
* Please consult your doctor, if this or any other problem occurs repeatedly
Other possible malfunctions and their elimination
If problems occur when using the device, the following points should be checked and if neces-
sary, the corresponding measures are to be taken:
Malfunction
The display remains empty
when the instrument is
switched on although the bat-
teries are in place.
11
Remedy
1. Check whether the batteries are installed with the correct
polarity and correct if necessary.
2. If the display is unusual, re-insert batteries or exchange
them.

Hide quick links:

Publicidad

Tabla de contenido
loading

Productos relacionados para Procheck WW1YB-3

Tabla de contenido