Error Codes
Error codes are shown on the main display for approx. 2 seconds, after which the program automatically returns to the weighing mode.
Display
No segments appear
on the display
h
l
e 01
e 02
e 09
e 10
e 11
e 22
e 30
Max. weighing capacity
is less than indicated
under "Specifications"
The weight readout
is obviously wrong
If any other errors occur, please contact your local Acculab Service Center.
22
Cause
No AC power is available
The AC adapter is not plugged in
Battery is dead
The load exceeds the balance/scale capacity
The weighing pan is not in place
Something is touching
the weighing pan
Display capacity exceeded:
Value to be output cannot
be shown on the display
Calibration parameter
requirement not met; e.g.:
– balance/scale not zeroed
– balance/scale is loaded
When gross value < zero, no tare
The [ZERO] key is blocked when
there is data in the second tare
memory (net-total) – only 1 tare
function can be used at a time
Value input is not allowed for
second tare memory
Weight is too light or there is
no sample on the balance/scale
Interface port for printer output is blocked
The balance/scale was
turned on without the
weighing pan in place
The balance/scale has not
been calibrated/adjusted
The balance/scale was not
tared before weighing
Solution
Check the AC power supply
Plug in the AC adapter
Replace the battery
Recharge the battery using an external charger
Unload the balance/scale
Place the weighing pan on the balance/scale
Move the object that is
touching the weighing pan
Decrease the weight on the balance/scale
Calibrate only when zero
is displayed
Press [ZERO] to zero the balance/scale
Unload the balance/scale
Tare the balance/scale
Press [CAL] to clear the tare memory and release
the tare key
Press [ZERO]
Increase the reference quantity
or sample amount
Contact your local Acculab Service Center
Place the weighing pan on the
balance/scale and press [ON/OFF] to
turn the balance/scale back on
Calibrate/adjust the balance/scale
Tare before weighing