POWER OFF
Press and hold for 3 seconds: power off
SENSORS
There are 5 sensors, one of them situated in the centre of the detection surface; it is visible
through the transparent casing. We recommend using this sensor to perform your tests on small
parts (fasteners, etc.).
The other 4 sensors are situated in the corners of a 1.5 cm square.
DETECTING MAGNETISM
Place the item on the detection zone (green)
IN CASE OF MAGNETIC REMANENCE
In addition to the value displayed in µT, up to 5 dots appear on the LCD screen, indicating by
their presence the magnetisation surface and by their size (bigger or smaller dot diameter) the
intensity of the magnetism between the various sensors.
The value displayed in µT will always be the highest value recorded by the 5 sensors.
If the LCD screen indicates the presence of magnetism without there being a magnetic item on
the detection surface, it is recommended to reset manually, by pressing the OFF button. (Effect
of the Earth's gravity).
DEMAGNETISING
Place the item on the demagnetisation zone (blue), and press the ON button once; the device
will perform its demagnetisation cycle. At the end of the cycle, please perform a check on the
detection part (green).
If remanence is detected, please repeat the demagnetisation process until you obtain a display
of 0 µT.
WARNING
Do not open the device; high voltage (> 400 V)!
The maximum value that the device can display is 999 µT, and the maximum value that the
sensors can withstand before deterioration is 100,000 µT.
Do no place any magnets on the device.
10