Front panel

F-333
RF beauty
instrument

Instructions of use

1. Connect the power supply via the rear panel.
2. Connect the accessories properly to the front/rear
panel as indicated in the sketch.
Accesory connection
Number
Accessorie/connection
1
Body electrode holder wire
2
Negative plate wire
3
Face electrode holder wire
4
Negative plate
5
Face electrodes
6
Body electrodes
7
"POWER" button
8
Power cord
3. Turn on the "POWER" button.
4. Press "SELECT" button to choose the desired
electrode holder (face or body). The display will show
the electrode holder that is working.
5. Put the negative plate in contact with a large
surface of the skin. It must be close to the treatment
zone.
Note: to avoid burning to the customer, never forget to
put this plate in contact with the skin.
State
select
Start/Pause
6. Choose the desired electrode according to the
treatment. Depending the treatment zone is needed to
choose an electrode of a proper size.
7. Apply enough gel or cream in the treatment zone.
8. Press "START/PAUSE" button to start the
treatment.
9. Choose the wave state according the customer
need. Is recommended firstly choose discontinuous
wave (for sensitive skins). Then, you can choose the
other wave states according to the customer
need/feelings and according to the discretion of the
beautician.
10. Apply the electrodes on the customer's skin
making constantly circles, without stopping of making
movements. Not keep the electrode without moving at
the same zone during a long period of time, the
customer could be burned.
11. Once the treatment has finished press
"START/PAUSE" button.