§ After you have finished the measurement for
calibrating and adjusting the temperature, an
acoustic signal (3 beep tones) is emitted.
Display: Analysis end(ed)
§ Press [Info] to display the measured values.
> Any values outside the tolerance limit are displayed
in "Info" (calibration). Press the c key to exit
the display.
§ Activate the calibration/adjustment and temperature
adjustment routine on the unit by pressing the u
key. This concludes the temperature adjustment
routine. If the final temperature shows 163°C, for
example, the temperature is correct.
§ Return to the main menu by pressing the c key;
this concludes the adjustment routine. A record is
printed out automatically if a printer is connected to
the unit.
1-Point Adjustment
You should select the 1-point adjustment for
samples that are usually measured at one tempera-
ture. In the 1-point adjustment mode, several
analyzers can be adjusted to the same drying tem-
perature. In the weighing mode, select one of the
three programs (P1, P2, P3). In the heating program
(standard drying), select and save the desired tem-
perature. The heating time is a factory setting that
cannot be changed.
13