10 FuctIOn sPecIFIcAtIOn
Information
The Pulse Oxygen Saturation (SpO
Pulse Rate (PR)
Pulse Intensity (bar-graph)
spO
Parameter specification
2
measuring range
accuracy
average value
Pulse Parameter specification
measuring range
accuracy
average pulse rate
safety type
Pulse Intensity
range
Display Mode
)
2-digit digital tft display
2
3-digit digital tft display
Bar-graph tft display
0%~100%, (the resolution is 1%).
70%~100%:±2%,Below 70% unspecified.
Calculate the average value in every 4 measure value. the deviation
between average value and true value does not exceed 1%.
30bpm~250bpm, (the resolution is 1bpm)
±2bpm or±2% (select larger)
moving calculate the average pulse rate every 4 cardio-beat,s cycle. the
deviation between average value and true value does not exceed 1%.
Interior Battery, BF type application part
Continuous bar-graph display, the higher display indicate the stronger
pulse.
31
English