INSTALLATION
The biological incubator can be placed either on a horizontal plane, which on a
metal surface through vertical fastening a magnet on the back of the unit.
Because of the characteristics of the device is important to follow certain
precautions:
The device should not be used in potentially explosive atmospheres.
Remove packaging, and check the status of the unit.
If any signs of damage to vehicles, alert the manufacturer.
Packaging must be preserved throughout the warranty period.
The manufacturer does not accept returns without original packaging.
Place the unit on a flat surface or on the autoclave cover through the
magnet fastener (fg.1 ‐ POS.D), away from heat sources (as bunsen,
vaporiere, ovens, etc....).
The unit must be placed at a height suitable to the user, we recommend a
height of 90 cm.
Place the equipment so that the led warning is clearly visible from the
operator.
Connect the power cord through the special connector to the post on the
autoclave.
Place with great care the device on the metal surface, avoiding violent shock
that, given the fragility of the magnet could result in a damage.
Given the intensity of the magnetic field of magnet avoid placing the device
in the nearness of electronic objects that could break or demagnatize
(mobile phones, watches, mp3 players, memory cards for data storage,
credit cards, etc....)
Place the device sideways in front area of autoclave, because in the back
could interfere with the functioning of autoclave.