Autoclaving - BrandTech Scientific Transferpette S Instrucciones De Manejo

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Autoclaving

The Transferpette
(121 °C / 250 °F) and 20 min. exposure time ( t
1.
Eject the pipette tip.
2.
Autoclave the complete pipette without
any further disassembling.
3.
Allow the Transferpette
cool and dry.
Note:
The effectiveness of the autoclaving must be
verified by the user. Maximum reliability is ob-
tained with vacuum sterilization. We recommend
the use of sterilization bags.
Attention:
Prior to autoclaving, the volume adjustment
must be set on an available numbered volume
(e.g., 11.25 or 11.26 but not between), with
the volume change brake set fully unlocked.
If the pipette is autoclaved frequently, the piston
should be oiled with the supplied silicone oil in
order to preserve smooth movement.
If necessary after sterilization, tighten the connec-
tion between the hand grip and the pipette shaft.
5 ml + 10 ml Filter Transferpette
A hydrophobic PE filter is used as a safeguard
against liquid entering the pipette.
Change the filter if it becomes wet or contaminated.
• Use a flat object such as a screwdriver.
• Remove the filter without damaging the tip cone.
Remove the filter before autoclaving!
The instrument can be operated without a filter.
30
®
S is completely autoclavable at 2 bar (30 psi)
®
S to completely
) according to DIN.
e
®
S

Publicidad

Tabla de contenido
loading

Productos relacionados para BrandTech Scientific Transferpette S

Tabla de contenido