5. Commissioning, operation
5.6

Emptying depot and depressurised gas compartment

The depot can be emptied only when there is measuring gas inside. If the depot is
already empty, the procedure is cancelled with an error message.
1. Press "Empty depot".
EN
2. Select the gas compartment for back pumping the measuring gas (only available
when the "back pumping" function is selected).
Measuring gas is pumped back.
Measurement is finished.
Connection 
The measuring gas is directly pumped back into the measured gas compartment.
Connection 
The measuring gas is pumped into a pressurised external gas compartment (e.g. gas cylin-
der).
The maximal pumping capacity of the analytic instrument is:
10 bar abs. (SF₆-Q-Analyser)
12 bar abs. (g³-Q-Analyser)
External container 
With the function activated, the measuring gas is emptied from an external container at
connection  as well.
20
WIKA operating instructions analytic instrument, model GA11