HEINE E4-USB Instrucciones De Uso página 7

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 12
The HEINE Charger is intended for installation in medical areas outside the patient environment
(at least 1.5 metres from the patient or patient support pursuant to IEC 60601-1, see figure).
Fig1: The above figure defines the patient environment (minimum radius around the patient).
Hygienic Reprocessing
Instructions on hygienic reprocessing must be adhered to, based on national standards, laws and guidelines.
Classification according to KRINKO: non-critical
Spaulding Classification USA: non-critical
Before cleaning, disconnect the device from the power source.
In the event of suspected contamination, carry out hygienic preparation of the instrument.
The described cleaning and disinfection measures do not replace the specific rules applicable
for the establishment.
HEINE Optotechnik only approves the cleansing agents and procedures named in these
instructions for use.
Cleaning and disinfection may only be carried out by personnel with sufficient hygienic knowledge.
Observe the instructions of the manufacturer of the reprocessing media.
Do not use spray or immersion disinfection, dripping wet or heavily foaming tissues, automated
or ultrasonic reprocessing.
Procedure
Clean and disinfect the device manually (wipe clean and wipe disinfect).
Recommended agents
Cleaning agent: Neodisher
Disinfectant agent: quaternary ammonia compounds (e.g. Microbac
Maintenance and Service
The instruments do not require maintenance or service.
General Warnings
Check the correct operation of the device before use! Do not use the device if there are visible
signs of damage.
Do not use the device in fire- or explosive risk area (e.g. oxygen saturated or anaesthetic environments)
Do not modify the device.
Use only original HEINE parts, spare parts, accessories and power sources.
Repairs shall only be carried out by qualified persons.
The device must not be placed near strong Magnetic fields, e.g. MRI units.
Do not use the power supply outdoors.
The device must not be used with appliance outlets of extension leads.
The power supply is only approved for the use with ungrounded loads (Class II).
MediClean
®
MED 112712
tissues)
®
2016-03-30
7/40

Publicidad

Tabla de contenido
loading

Tabla de contenido