Adjustment Of Zero Point / Span - WIKA GCS-1 Manual De Instrucciones

Gas cylinder scale model gcs-1
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

7. Adjustment of zero point / span

7. Adjustment of zero point / span
Preparation
In order to access the zero point and span potentiometers, unscrew the protective
cap using a screwdriver (see figure A).
Introduce a reference mass which is at least 3 x as accurate as the given accuracy.
Zero point
Remove the load from the gas cylinder scale.
By adjusting the signal to 4 mA, the zero point can be set (see figure B).
Span
Set the span by loading the scale with the appropriate reference mass and adjusting
the signal to 20 mA.
Check the zero point.
If the zero point is not right, repeat the procedure.
Screw the sealing cap with seal back in, in order to close the potentiometer entry.
A
B
After the adjustment, check that the system is functioning correctly.
Recommended recalibration cycle: 1 year
WIKA operating instructions gas cylinder scale model GCS-1
GB
15

Publicidad

Tabla de contenido
loading

Tabla de contenido