RANGES AND SPECIFICATIONS
Test Factor
Alkalinity, Total
Biguanide
Biguanide Shock
Borate
Bromine (DPD)
Chlorine, Free (DPD)
Chlorine, Combined
Chlorine, Total (DPD)
Test results that are out of range of the reagent system wil be RED.
not be accurate. Go SETTINGS>Other Settings to turn the Ranges Enabled feature OFF.
Copper concentrations over 4 ppm will result in low total hardness results.
Instrument Type
Wavelengths
(interference fi lters)
Display
Wavelength Accuracy
Wavelength Bandwidth
Photometric Range
Photometric Precision
Photometric Accuracy
Sample Chamber
Light Source
Detectors
Pre-Programmed Tests
Languages
Temperature
Operation Humidity Range
Communication
Calibration
Firmware
Software
Power Requirements
Battery Type
Range
0–250 ppm
0–70 ppm
0–250 ppm
0–80 ppm
0.00–33.00 ppm
0.00–15.00 ppm
0,00–15.00 ppm
0.00–15.00 ppm
Centrifugal Fluidics Photometer
390 nm, 428 nm, 470 nm, 525 nm, 568 nm, 635 nm
Color Capacitive Touchscreen,
3.5 in, 320 x 240 pixel resolution
±2 nm
10 typical
-2 to 2 AU
±0.01 AU at 1.0 AU
±0.01 AU at 1.0 AU
Accepts prefi lled disk
6 LEDS
6 silicon photodiodes
Yes, with automatic wavelength selection
English, French, Spanish, German, Dutch, Swedish, Portuguese,
Italian, Chinese, Turkish
Operation: 0-50 °C; storage – 40-60 °C
0- 90 % RH, non-condensing
USB-C, Bluetooth low energy technology (BLE)
Factory set, fi eld calibration via internet connection
Internet updateable (New Test, New Test Calibrations, etc.)
Android & iOS Apps, DataMate
USB wall adapter, USB computer connection or internal lithium ion
rechargeable battery
Lithium ion
Test Factor
Copper
Cyanuric Acid
Hardness, Calcium
Iron
pH
Phosphate
Salt
®
Web, DataMate
16
Range
0.0–3.0 ppm
5–150 ppm
0–800 ppm
0.0–3.0 ppm
6.4–8.6
0–2000 ppb
0–5000 ppm
RED
test results may
®
10