Installation and start-up

Room temperature

The climate class is stated on the
nameplate (see page 10). This specifies the
optimum room temperature.
SP = -82
+16 ºC to +25 ºC
Ventilation
It is important that the appliance be
well ventilated and that air can circulate
unhindered above, below and around it.
Figure 2 illustrates how the necessary air
circulation can be ensured.
Placement
For safety and operational reasons, the
appliance must not be installed outdoors.
The appliance should be placed on a
stable, level surface in a dry, well-ventilated
room (max. 75% relative air humidity).
Never place the appliance close to sources
of heat such as radiators, and avoid placing
it in direct sunlight.
SP = -86
+16 ºC to +20 ºC
There must be at least 35 mm clearance
between the base of the appliance and the
floor.
IMPORTANT! There must be free access to
the ventilation grille (fig. 3).
5
GB
fig. 2
fig. 3