Expected Service Life - arjo 22000 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 10
Applying the Limb Flites

Expected service life

Storing and Shelf life
Safe Working Load (SWL)
Product care & Maintenance
Service & Maintenance
Customer support/Trouble shooting
1. Fold the Limb Flites (to avoid straps touching the skin). Place it below
the patient limb in the desired position, and carefully pull the straps to
each side.
2. Check patient position and limb sling position carefully before lifting.
3. Place the loop or clip straps in the hangerbar according to the IFU for
the lifter
4. Carefully raise the hangerbar and perform the lifting, while keeping
eye on the patient's position.
The service life is expected to be 56 performed lifting sessions, however
this will be subject to the actual conditions during use.
The Limb Flites are disposable and designed for single patient use and
should be disposed after patient hospital stay or when soiled or subjected
to liquid.
The Limb Flites should be stored in a dry and clean environment and the
shelf life is 5 years in original packaging. Store in temperatures between
0°-40° C (+32°-104°F)
Max 70 kg / 154 lbs.
Always follow the instructions on the product label and check the sling
for deteriorations.
Mark the sling with the specific patient name and keep in dry and
clean environment.
This is a patient specific disposable sling, that must be disposed after
the patient has been discharged.
The Limb Flites shall be checked by authorized personnel every 6 months
according to ISO10535.
Please refer to the IFU for the lifting equipment used.
If in doubt, please contact your dealer who supplied the product
UK

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

22001

Tabla de contenido