VWR Thermal Shake lite Manual Del Usuario página 12

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

OPERATION GUIDE
This chapter gives a through instruction to the temperature set up, vibration time set up,
vibration speed set up of this instrument. It also explains how to correct the discrepancy
between the actual temperature and displayed temperature, and how to set up the
buzzer.
1
Pre-check before turn on power
Before turning on power, check and confirm the following items first:
1. Unit is on stable and level surface;
2. The voltage should comply with the spec; (Refer to Chapter 2 for the power
supply spec);
3. The power plug should be tightly plugged into the outlet;
4. The ground connection should be reliable;
5. The module connection should be reliable.
If the display is abnormal when the power is turned on, please shut
down immediately and contact the supplier.
2
Set up of temperature, vibration time and vibration speed.
a) Approximately 5 seconds after the power is
turned on, the monitor will display a temperature of
28.5, which is the instant temperature of the module.
The display of "TEMP 100.0" is the set up
temperature of last run, the display of "TIME 80:00" is
the set up vibration time of last run, and the display of
"SPEED 1000" is the vibration speed of last run.
At same time, the device will automatically heat or
cool, to achieve the setting temp. point.
9
5
Section
TEMP
TIME
SPEE

Publicidad

Tabla de contenido
loading

Tabla de contenido