6.2.11 Info; Web Browser - Sartorius stedim BioWelder Manual De Funcionamiento

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

26
Operation

6.2.11 Info

In this menu the current status is displayed according to following table:
Page 1
Item
Meaning
IP address
Currently used IP address
DHCP
Yes or no, depending on whether DHCP is used
Serial number
Serial number of the device
Page 2
Item
Meaning
Firmware
Firmware version
File system
File system version
Weld Cycles
Number of welding cycles which the device has executed
so far
Tube Holder Cycles
9999 (fixed number)
- (no tubeholder inserted)
Wear
Equipment abrasion indicator . If the wearing down of one or
more parts has reached ≥100%, information will appear on
the operator screen . Refer to Chapter 5 .5 Maintenance .
6.3

Web browser

The Operator and Supervisor users can perform their tasks equally with the touch
screen or the Web browser .
Advantages of working with the Web browser:
No additional software needs to be installed .
®
Several BioWelder
TC devices can be operated simultaneously over a Web browser .
Operation via mobile devices such as laptop, tablet PCs, etc . is possible .
All known browsers on different platforms are supported .
The prerequisites for working via Web browser are:
A computer with Web browser and a network cable are available .
The network configuration in Chapter 6 .3 .5 General settings was correctly done .
The following Web browsers are recommended:
Internet Explorer from version 8
Firefox from version 7
Google Chrome from version 16
Greyed out menu items cannot be chosen as the prerequisites have not
been met.