Setting Ramp Time - ResMed S8 Compact System Manual De Instrucciones

Ocultar thumbs Ver también para S8 Compact System:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

S
R
ETTING
If the patient has difficulty falling asleep with full pressure, select a ramp time. The
pressure will slowly increase to full operating pressure over the selected ramp time.
The Ramp Time screen is displayed after the standby (Ramp) screen, and is also
available in the Clinical Menu. The clinician also sets a Maximum Ramp time; the patient
may select any value up to the set maximum.
When ramping begins the LCD will display the word RAMPING with the current
pressure indicated as dashes. Table 3 describes the range of set pressures that the
dashes represent.
Table 3: Pressure range indicators on S8 LCD
No. of dashes
2
3
4
5
6
7
8
9
10
20
T
AMP
IME
Pressure range (cm H
Set Pressure ≥ 4.0 and ≤ 5.0
Set Pressure > 5.0 and ≤ 7.0
Set Pressure > 7.0 and ≤ 9.0
Set Pressure > 9.0 and ≤ 11.0
Set Pressure > 11.0 and ≤ 13.0
Set Pressure > 13.0 and ≤ 15.0
Set Pressure > 15.0 and ≤ 17.0
Set Pressure > 17.0 and ≤ 19.0
Set Pressure > 19.0 and ≤ 20.0
0)
2

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido