Instructions for use (rev 120121)
4.2.

INSTALLATION AND USE

Note: for marks, see chapter A – equipment description.
1.
Open the pivoting cover (e).
2.
Unscrew the tank's cap (c).
3.
Remove the tank (b) of equipment and fill-in with the disinfectant liquid choose (use a suitable funnel
if necessary). Attention to do not exceed the maximum tank's volume (2 litres).
IMPORTANT: not exceed the maximum tank's volume (2 litres).
4.
Replace the tank in the equipment (in the space provided).
5.
Screw the tank's cap (c) and close the pivoting cover (e).
6.
Make sure that the mains voltage complies with the voltage indicated on the rating plate, connect the
power cord (j) to a 10/16 A grounded power socket.
7.
Put the equipment in the middle of the room to be treated, and keeping a distance of at least 1 meter
from anything furniture and wall.
8.
Activate the ON/OFF switch (h).
9.
Make an estimate of the room's volume in m
volume to be treated.
IMPORTANT: turn the volume selector clockwise direction. Never selector manually counter-
clockwise direction: may cause the deterioration of the selector.
10. The spraying starts 60 seconds after the volume is selected, which allows the user to leave the room
before start the product's spraying.
11. The spraying continues for the programmed time (according to selected volume), then stops
automatically at the end of the cycle.
If the tank contains liquid from a previous use, make sure that it contains enough liquid to perform the
operation. If necessary, add the required amount of the same product (do not exceed the maximum volume
of 2L).
IMPORTANT: do not operate equipment without liquid in the tank.
This may result in malfunction of the unit and damage to certain components.
▪ CAUTION: mixing of disinfectants is not recommended - dangerous mixtures may form. Observe the
manufacturer disinfecting's instructions. Consult the product data sheet and the material safety data
sheet (MSDS).
▪ Do not use flammable, corrosive, toxic liquids.
▪ Before use, consult the disinfectant product's technical data sheet to check the compatibility with the
floor/wall covering and the furniture/materials of the room to be treated (for example, certain liquids
do not recommend using them on copper, copper alloys or light metals such as aluminium). If in doubt, it
is recommended that you test a patch of the surface to be cleaned.
▪ Room's temperature and humidity conditions can cause the aerosols of the disinfectant liquid to
condense. It is recommended to use Aseptojet S in a room at ambient temperature (15°C - 30°C).
UGIN DENTAIRE
3
. Set the volume selector (f) according to the room's
ASEPTOJET S
15