Reading test data
• Read the test values on the display (4)
• By pressing the key
The position display reverts back to the absolute jaw spacing.
• By pressing the key
Remove test piece
• Relieve clamping jaws (11) with locking lever (9) and remove welding sample.
• The tool is ready for further testing.
Ready to transport
• Push locking lever (9) down until to the limit.
• Unplug tool from the line/mains.
• Unplug cable from socket (1) und put it into the storage case.
• Close the storage case.
Change of clamping jaws
• Flexible jaws
– Fasten adjustment screw (10) against the clamping jaw (11).
– Unscrew the cocking-lever shaft (22) from clamping jaw (11)
with a 8 mm spanner.
– Remove clamping jaw (11).
• Fixed clamping jaws
– Unscrew set screw (16) with 4 mm hexagonal box spanner.
– Pull off clamping jaw (11).
• Assemble clamping jaws in reverse order.
– Watch the alignment of clamping jaws (11).
22
, the sliding carriage (12) will return to its programmed starting position.
–
‹‹
› again, the test data are cancelled and a new tensile test can be started.