Connections; System Functions - HACH LANGE si629 C Manual

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

3. Installation
Details
To open the transparent front door, press with two
fingers the part marked with ( ) in the scheme.

3.1.2. Connections

When making any wiring connections to the instrument, any warnings and notes found throughout the
manual must be adhered to. For more safety information see "Safety information" on page 5.
DANGER
Always disconnect power to the instrument when making any electrical connections.

System functions

DANGER
High-voltage wiring for the instrument is protected by a protective casing. The protective casing
must remain in place unless a qualified installation technician is installing wiring for power.
8
si629 C_HACH LANGE_Ed 0510.indd ANG:8
si629 C_HACH LANGE_Ed 0510.indd ANG:8
Cable glands.
On the lower part of the instrument there are 5
cable glands, 2 PG 7 and 3 PG 9.
1,2
Voltage 230 or 24 VCA ± 10%.
(115 VCA upon request).
3
Ground connections, see
notes (
page 9).
4,5
Current output 4-20 mA (EC).
6,7
Input for temperature probe.
8
Cell cable shielding
9 to 12 Entrada para célula de 2 polos.
13,14
Limit 1 relay contacts.
15,16
Alarm relay contacts.
17 to 19 Limit 2 or Cleaning relay contacts.
PG 7 (2 units)
PG 9 (3 units)
Important
20/5/10 12:06:50
20/5/10 12:06:50

Publicidad

Tabla de contenido
loading

Tabla de contenido