Troubleshooting; Cleaning; Service And Maintenance - Maxtec MaxFLO2 Manual De Instrucciones E Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 93

4.0 TROUBLESHOOTING

Problem: Gas leaking past flow control valve even though valve is closed.
Potential causes and solutions:
• Flow valve knob is over-tightened or the valve's internal stem may not be seated
correctly. Turn the valve counter-clockwise to allow some flow then re-close the valve
without undue force.
• Flow valve is contaminated or internal o-rings are worn or damaged. Contact your
Maxtec distributor or Maxtec for repair.
Problem: Float ball is not stable and bobbles up and down.
Potential causes and solutions:
• Flow rate is outside flowmeter specifications. Reduce flow rate. (Occasional and
intermittent bobbling is normal at flow rates between 80% and 100% of full scale.)
Problem: Flow exiting the MaxFLO
Potential causes and solutions:
• Excessive backpressure is reducing sensitivity of flowmeters. Reduce the total flow
rate and/or use less restrictive patient circuitry. Note: The FiO
Table 1 (pg.3) will not be affected.
• Float ball is contaminated. Contact your Maxtec distributor or Maxtec for repair.
Problem: Analyzer/Monitor diverter is difficult to insert or does not remain inserted in
Mixed Gas Analysis Port.
Potential causes and solutions:
• Diverter is the wrong size. Use diverter supplied by Maxtec, part number
R110P10-001 as noted in section 10.2 of this operating manual.

5.0 CLEANING

Use caution to prevent any fluid from entering the MaxFLO
of the MaxFLO
2
before re-use.
DO NOT soak or immerse the MaxFLO
DO NOT allow any fluid to enter the MaxFLO
DO NOT autoclave or expose the MaxFLO
DO NOT use any solvent based cleaners on the MaxFLO

6.0 SERVICE AND MAINTENANCE

Maxtec recommends the performance check listed in section 3 of this operating
manual be conducted prior to placing the MaxFLO
thereafter. If the MaxFLO
discontinue use of the device pending service by trained technicians or contact your
Maxtec distributor or Maxtec at:
2305 South 1070 West, Salt Lake City, UT 84119 (801) 266-5300 or (800) 748-5355
The MaxFLO
's flow adjustment valves should be replaced as needed with R219P33
2
from section 10.2 of this operating manual. The full MaxFLO
6
2
in accordance with institutional cleaning protocols. Dry completely
does not function as described in the performance check,
2
866.4.Maxtec
is higher than the cumulative flow of both flowmeters.
in fluid.
2
.
2
to ethylene oxide sterilization.
2
into clinical use and periodically
2
www.maxtec.com
readings from
2
. Clean the exterior surface
2
or its' labels.
2
device should be
2

Publicidad

Tabla de contenido
loading

Tabla de contenido