CLEANING
CAUTION
DO NOT steam autoclave.
DO NOT immerse the Air-Oxygen Blender into any liquid.
DO NOT use any strong solvent or abrasive cleaners.
DO NOT gas sterilize with (EtO) Ethylene Oxide.
DO NOT clean with aromatic hydrocarbons.
DO NOT allow the liquid to penetrate the device.
DO NOT gas or heat sterilize.
1. Disconnect all gas connections and equipment before cleaning.
2. Close Bleed Valve to ensure O-Ring is not exposed.
3. Clean exterior surfaces using Super Sani-Cloth germicidal disposable wipes. Remove all visible
contamination from the external surfaces of the device and its accessories. Be sure to closely
inspect and remove contamination from seams and recesses on the device that may trap
contaminants. Wipe with clean paper towel to remove debris.
4. After all visible contamination is removed; use a second germicidal wipe to thoroughly wet
the surfaces of the device and accessories. Allow to remain wet for 4 minutes. Use additional
wipes if needed to assure surfaces are wetted continuously for 4 minutes.
5. Allow device to air dry completely.
6. Visually inspect the device for visible contamination. Repeat cleaning/disinfection process if
necessary.
MAINTENANCE
The following maintenance on the Air-Oxygen Blender must be performed by a trained service
technician:
◆ • The alarm should be tested prior to being placed into
clinical service and periodically there after.
◆ • Every year conduct the Operational Verification Procedure (OVP).
* A detailed description of the OVP tests can be found in the Blender Service
Manual (P/N R203M10), available on the Internet; www.maxtec.com.
◆ • Every 2 years the Air-Oxygen Blender should be serviced.
◆ • Refer to the Air-Oxygen Blender Service Manual (P/N R203M10) for
complete details regarding further maintenance and testing.