1
OIL
30233040B PBK9- / K...(x-T4)-series
4.2 Cleaning the interior
The load plate has to be removed in order to clean the load cell.
NOTICE
Damage to the load cell due to improper handling.
▲ Never touch, direct compressed air against or spray the rubber membrane (1) of
the load cell.
1. Remove load plate, for PBK9..-CC / KCC...(x-T4) see section 4.4.
2. Blow out dirt or flush out with a medium-power water jet (< 2 bar).
3. Replace load plate, for PBK9..-CC / KCC...(x-T4) see section 4.4.
4.3 Subsequent treatment
In order to protect the scale, carry out the following subsequent treatment:
DANGER
Danger of injury due to failure of the pneumatic springs.
▲ With raisable load plates do not oil the piston rods of the pneumatic springs.
➜ Rinse the scale with clear water and remove cleaning agent completely.
➜ Dry off the scale with a lint-free cloth.
➜ With stainless steel scales, treat the inside and outside with an oil suitable for
foodstuffs. With raisable load plates and recessed scales ensure that all moving parts
and hinges are treated with oil, too.
11
Maintenance