Annex
Annex
Guidance and manufacturer´s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and should
only beused in such environments:
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fl uctuations / fl icker
emissions IEC 61000-3-3
Guidance and manufacturer´s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and should
only beused in such environments:
IEC 60601
Immunity test
– test level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Power frequency
(50/60 Hz)
3 A/m
magnetic fi eld
IEC 61000-4-8
Recommended separation distances between portable and mobile RF communication
The device is intended for use in an electromagnetic environment where radiated RF
disturbancesare under control. User can help prevent electromagnetic interference by keeping
the device at a minimum distance from portable and mobile RF communications equipment
(transmitters). Below table details the maximum output power of transmitter:
Rated maximum
Seperation distance according to frequency of transmitter / m
output power
of transmitters
in Watt
150 kHz to 80 MHz
d = 1.2 √P
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
For transmitters rated at a maximum output power not listed above, the recommended separa-
tion distance d in metres (m) can be estimated using the equation applicable to the frequency
of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
At 80 MHz and 800 MHz, the separation distance for the higher frequency
NOTE 1:
range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propaga-
tion is affected by absorption and refl ection from structures, objects and
people.
Compliance
Electromagnetic environment - guidance
RF energy is used only to maintain device's
operation. Therefore, its RF emissions are so
Group 1
low that it's not likely to cause any interference
in nearby electronic equipment.
Class B
The device is suitable for use in all establish-
ments, including domesticestablishments,
Class A
and those directly connected to the public
low-voltage power supply networkthat supplies
buildings used for domestic purposes.
complies
Compliance
Electromagnetic environment
level
- guidance
Floors should be wood, concrete or
± 6 kV contact
ceramic tile. If fl oors are covered with
± 8 kV air
synthetic material, the relative humidity
should be at least 30 %.
Power frequency magnetic fi elds
should be at levels characteristic of a
3 A/m
typical location in a typicalcommercial
or hospital environment.
80 MHz to 800 MHz
d = 1.2 √P
0.12
0.38
1.2
3.8
12
45
800 MHz to 2.5 GHz
d = 2.3 √P
0.23
0.73
2.3
7.3
23
EN