7 Trouble-shooting

The following table shows explanations of the error messages and tips on how to
avoid incorrect measurements.
The most commonly encountered problems are caused by:
The test strips or cells not being inserted correctly.
Cells being dirty or strongly scratched.
The reaction time of the strip not being right.
Incorrect use of the tests (e. g. strips / tests not properly stored, pH range
not cor
rectly adjusted).
Problem
Cause
• No display
• Batteries not
en
• Batteries are drained
• Display LO
• Concentration is
• Display HI
• Concentration is hig-
• Display - - -
• No calibration pro-
|
28
Solution
• Insert the batteries
inserted properly
correctly
• Use new batteries
• State the result is
lower than the meas-
smaller than the
urement range of the
lowest measurement
test
value
• Dilute the sample; do
her than the measu-
not forget to consider
rement range of the
the factor for dilution
test
when reporting the
result!
• Calibrate instrument
grammed
Problem
Cause
• Display OPT
• Optics or strip adap-
ter is dirty
• Display ERR
• Optical error
• Strip adapter is
assembled incorrectly
• Display E-1
• Surrounding is too
bright
• Strip adapter is
assembled incorrectly
• Display E-2
• Result is not plausible
Solution
• Clean optics and
adapter and switch
the RQflex® plus 10
on and then off
again; recalibrate, see
page 26
• Clean optics and
adapter; recalibrate
instrument, see page
26; if necessary, call
customer service
• Repeat measurement
in a darker place;
en
if necessary, close
the lid of the instru-
ment before switching
on
• Repeat measurement
|
29