will be displayed as a negative number.
3. You may now fill the container with the quantity to
be weighed. The display will show the net weight of
the contents.
4. To return to normal weight display (gross weight),
remove any items from the weighing platform and
press the TARE key again to re-zero the scale.
VII.
Calibration
Calibration and adjustment are performed at the factory.
However, the weighing range can shift slightly during
shipping. Also, due to changes in local acceleration of gravity
around the world, adjustment is recommended when moving
the scale from one location to another. To check calibration:
1. Place the required calibration weight on the scale
and note the displayed value.
2. If the displayed value is not within acceptable
tolerance (see chart below), adjustment is required.
Model
AWS-70
AWS-100
AWS-100-CAL
AWS-201
AWS-250
AWS-600
AWS-1KG
Accept-
Calibration
able
Tolerance
(±)
0.02g
0.02g
0.02g
0.02g
100g & 200g
0.2g
0.2g
0.2g
Recommended
Weight
Accuracy Class
50g
100g
100g
200g
500g
500g
OIML M2
OIML M2
OIML M2
OIML M2
OIML M2
OIML M2
OIML M2