Descargar Imprimir esta página

Kübler CODIX 551 Manual De Instrucciones página 40

Publicidad

Idiomas disponibles

Idiomas disponibles

5 5 . . 3 3 . . 1 1 . . 5 5 C C o o r r r r e e c c t t i i o o n n v v a a l l u u e e
Entry of a correction (offset) value, by which the measured value will be adjusted. The
value will be added to the measured value. Positive as well as negative correction values
are possible. The entry is always carried out to one decimal point.
Menu <–> Select
press the
Assuming the measured value is 28.45 and the correction value is –1.5, the display
will indicate 26.95 .
5 5 . . 3 3 . . 2 2 R R e e s s i i s s t t a a n n c c e e M M e e a a s s u u r r e e m m e e n n t t
Menu <–> Selection
press the
Also suitable for non-linear resistances
5 5 . . 3 3 . . 2 2 1 1 S S e e l l e e c c t t M M e e a a s s u u r r i i n n g g M M o o d d e e
Menu <–> Selection
press the
4. electrical connections
Page 14
Set digit,
Select digit
keys to accept the selection
keys to accept the selection
2 wire
3 wire
4 wire
keys to accept the selection
9
www.kuebler.com

Publicidad

loading