Consumable material and options
For applications in the dental area:
If necessary, use a yellow filter in order to avoid the composite filling material hardening,
as this blocks UV radiation.
The plug-on polarisation filter P2 enables a differentiation and low-reflex examination of
tissue surfaces due to polarized light. Only in combination with ML4 LED light head or
LED LoupeLight and HR 2.5x binocular loups, ready to use. The polarisation filter P2 is
not a replacement for a yellow filter as it may harden composite filling material.
The setup and operation of the HEINE power sources are described in a separate
instruction document.
Hygienic Reprocessing
Instructions on hygienic reprocessing must be adhered to, based on national standards,
laws and guidelines.
Classification according to KRINKO: noncritical
Spaulding Classification USA: noncritical
Allow the device to cool down before reprocessing.
Before cleaning, disconnect the device from the power source.
Before you plug the device back in, dry the plugs of the LED LoupeLight and the
LED MicroLight.
In the event of suspected contamination, carry out hygienic preparation of the
instrument.
The device and its accessories are not sterile. Avoid touching with sterile hands,
areas or instruments.
The described cleaning and disinfection measures do not replace the specific rules
applicable for the establishment.
HEINE Optotechnik only approves the resources and procedures named in these
instructions for use.
Cleaning and disinfection may only be carried out by personnel with sufficient
hygienic knowledge.
Observe the instructions of the manufacturer of the reprocessing media.
Do not use any spray disinfectant, or any dripping-wet or heavily-foaming cloths.
Do not prepare mechanically.
For the polarisation filter P2, do not use any alcoholic preparing agents.
Procedure
Clean and disinfect the head lights, the options and the headbands manually
(wipe-cleaning and wipe-disinfecting). Clean the leather cushion of the headbands with
a detergent cloth. Conditioning residues on optics may be removed using a low-bacterial,
dry, micro-fibre cloth.
Recommended agents:
Cleaning agent: Neodisher
Disinfectant agent: quaternary ammonia compounds (e.g. Microbac
Please follow the cleaning notes for the S-Frame in the identically named instructions
for use.
Number of preparation cycles
Polarisation filter P2 may be re-prepared up to 50 times.
Maintenance and Service
The instruments do not require maintenance or service.
The expected operational life of the HEINE head light is 5 years.
MediClean
®
med 0714
2016-06-30
tissues)
®
13/76