Altitude <2000m
Operating temperature: 0 ~40°C (<80% RH, non condensing)
Storage temperature: -10 ~ 60°C (<70% RH, battery removed)
Max. Voltage between terminals and earth ground: 600V RMS
Operating principle: dual slope integration
Sample rate: 2 times/sec for digital data
20 times/sec for analog bar
Display: 3¾ digits LCD display with max. reading 3999
42 segments fast analog bar display.
Automatic indication of functions and symbols.
Range selection: autorange and manual range.
Over range indication: LCD will display "OL" (AC/DC range).
Polarity indication: "-" displayed automatically.
Law opening diameter: cables Ø28mm
Maximum conductor size: Ø28mm
Low battery indication: The
operation range.
Auto power off time: If there is no key operation for 30 minutes, the meter will power
itself off to save battery consumption. This function can be disabled by press and hold
the ZERO key then power the meter on.
Power source: DC 1,5V x 2 Size AAA
Dimensions: 194(L) x 72(W) x 35(H) mm
Weight: 210g approx. (batteries included)
Accessories: User´s manual, test leads, carry case.
Measurment specifications
-
Position the conductor within the jaws at the intersection of the indicated marks as
much as possible in order to meet this meter´s accuracy specifications.
-
If the conductor is positioned elsewhere within the jaws, the max. additional error
resulted is 1,5%.
Accuracy: ± (% of reading + number of digits) at 18°C to 28°C (64°F to 82°F) with relative
humidity to 80%.
DC current (autorange)
Range
40A
400A
Maximum input current: 500A DC up to 60 seconds
is displayed when the battery is under the proper
Resolution
0,01A
0,1A
15
Accuracy
± 2,5% +5