PERIODICAL MAINTENANCE CHECKS

The SUPER VEGA suction equipment does not need maintenance or lubrication.
It is necessary to check functioning and instrument before every use. Unpack the instrument and always
check integrity of plastic parts and feeding cable, they might have been damaged during previous use.
Connect cable to electrical network and turn switch on. Close the aspirator outlet with your finger and with
suction regulator in maximum vacuum position check that the vacuum indicators reaches
-80kPa (-0.80 bar) minimum. Rotate the knob from right to left and check the aspiration regulating control.
The vacuum indicator should go down -40kPa (-0.40 bar).
Verify that loud noises are not present, these can indicate wrong functioning.
Two protection fuses (F 1 x 1.6 A 250V) reachable from exterior and situated in the plug protects the
instrument. For fuses replacing, always the type and the range.
Type of fault
1. No aspiration
2. No aspiration
3. The float doesn't close
4. The float doesn't close
5. Low suction
6. No aspiration due to
flow leakage of mucus
Faults 1 - 2 - 3 - 4 - 5 - 6
If the overfill security system it's activated, don't proceed with the liquid aspiration.
If the overfill security system doesn't work there are two cases:
1° case – If the overfill security system doesn't work the aspiration will be stopped by the bacteriological
filter who avoid the liquid penetration inside the device.
2° case – If both the security system doesn't work, there is the possibility that liquid comes inside the
device, in this case return the device to GIMA technical service.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING,
PLEASE CONTACT GIMA TECHNICAL SERVICE.
GIMA DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL SERVICE
CHECKING, APPEARS TO BE TAMPERED.
IMPORTANT INFORMATION FOR CORRECT DISPOSAL OF THE PRODUCT IN ACCORDANCE WITH
At the end of its working life, the product must not be disposed of as urban waste.
It must be taken a special local authority differentiated waste collection or to a dealer providing this
service. Disposing of a household appliance separately avoids possible negative consequences for the
environment and health deriving from inappropriate disposal and enables the constituent materials to be
recovered to obtain significant saving in energy and resources. As a reminder of the need to dispose of
separately the product is marked with a crossed-out wheeled dustbin.
Cause
Jar Cap badly screwed down
Lid seal not in its seat
If the cap has been washed,
ensure that the float is not partially
detached
The float it's covered by dirty
material
Foam inside the jar
Filter blocked
None of the remedies has
achieved the desired results
EC DIRECTIVE 2002/96/EC:
12
Remedy
Unscrewed the cap, then rescrew
it correctly
Unscrew the cap and insert the
seal properly in its seat
Insert the float into it's place
Unscrewed the cap, leave the and
put in on autoclave
Fill the jar to 1/3 full of ordinary
water
Replace filter
Contact the seller or GIMA After-
sales Assistance Service