Setting up the Instrument

The R&S ZN-Z15 is designed for indoor use and can be set up on a level surface
in a horizontal position. It is not designed for rack mounting.
Potential damage caused by electrostatic discharge
Electrical discharges can damage components of the instrument or a con-
nected instrument under test.
For this reason, the instrument must only be operated at a workplace that is
protected against electrostatic discharge.
The following methods can be used separately or in combination to protect
against electrostatic discharges:
● Protective wrist strap with a ground conductor
● Conductive floor covering combined with use of a heel grounder
EMI Suppression
To suppress electromagnetic interference (EMI), the instrument may only
be operated when it is closed and all covers are in place. The instrument's
EMC rating is specified on the data sheet.
To prevent influences that cause disturbances, the following conditions
must be adhered to:
● Use suitable double-shielded cables.
● Don't use any USB cables that are longer than 1 m.
● Only use USB devices that maintain the permitted EMC values.
● For the LAN connection (Ethernet), use shielded CAT6 or CAT7 RJ-45
cables (F/STP or S/STP)
User Manual 1177.5479.02 ─ 02
Putting the Instrument into Operation
Preparing the Instrument
11