Air Polishing Using The Jetshield ™ Accessory - DENTSPLY Cavitron PROPHY-JET Manual Del Usario

Ocultar thumbs Ver también para Cavitron PROPHY-JET:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

To adjust the flow of powder:
• Adjust the powder flow rate by rotating the powder
control knob on the front panel.
• For heavy stain removal, rotate fully clockwise.
• For light stain removal, rotate fully counter clockwise.
• The control can be set at any position as necessary.
• The view window at the center of the pointer lets you
observe the powder flow (a white fluffing of powder)
during operation. If no flow is seen, check for clogging
or add cleaning powder.
8.3 Air Polishing using
the JetShield
Aerosol reduction device for use with air polishing systems.
JetShield
assembly and accessories—see Fig. 1. The disposable
JetShield
assembly is designed for single use; all other
components are reusable and are compatible with standard
clinic equipment.
Figure 1
A. Ejector tube
F. Connector
E. Ejector tube clip
Precautions:
The JetShield
itself is designed for single use only. Use of
chemical or heat sterilization will lead to poor performance
and malfunction of product.
The reusable components are wipe-disinfectable, using clean
cloths and appropriate non-immersion type disinfectant. Wipe
once with disinfectant to remove gross debris, then use fresh
cloth and more disinfectant to clean surface. Dry with clean
paper towel or air.
Accessory
B. JetShield
assembly
C. Connector
D. Handpiece clip
Step-by-Step Instructions:
1. Remove entire assembly from package.
2. Insert air polishing nozzle through JetShield
and seat shoulder flush on nozzle. See Figure 2.
3. Insert connector on JetShield
the flexible ejector tube.
4. Insert connector of ejector tube into operator saliva ejector.
Figure 2
5. Place ejector tube clip around ejector tube and handpiece
cable such that cables drape together for patient and clinician
comfort.
6. (Optional) Place handpiece clip along handpiece, close to
handpiece/cable connection, such that clip is holding the
JetShield
7. Verify that slurry exits from the JetShield
unit is activated. Adjust saliva ejector as necessary.
8. Proceed with routine air polishing, applying cup to tooth
so cup flares slightly. Pivot nozzle inside cup, adapting to
mesial and distal surfaces. Apply 2 seconds of spray for each
segment of tooth, releasing foot control each time. If spray
escapes, readjust cup to maintain flat contact with tooth sur-
face, adjusting angle by pivoting nozzle inside of handpiece.
After completion of each procedure, remove JetShield
and dispose of JetShield
Wipe-disinfect all other components before next use.
Section 8: Techniques for Use con't
assembly into
1. Insert air polishing nozzle
2. Seat shoulder flush on nozzle
3. Incorrect positioning
evacuation tube in place along handpiece.
according to applicable regulations.
cup
when air polishing
assembly
ENGLISH
9

Publicidad

Tabla de contenido
loading

Tabla de contenido