Cleaning and Disinfecting
Warning: Cleaning and disinfecting are separate processes. Cleaning must precede disinfection.
Before use by a different individual: clean, disinfect, and check product for proper functioning.
Warning: Follow all manufacturer bleach container and germicidal disinfectant safety guidelines.
Cover Care: Remove hook fastener pieces from bottom of cover prior to cleaning and disinfecting. Wash with
like colors.
To Clean the Cover: Remove the cover. Machine wash in cold water (30°C) with mild
detergent, gentle cycle, or wipe clean with neutral detergent and warm water. Tumble dry low.
To Disinfect the Cover: Hand wash in cold water (30°C) using
1 part household liquid bleach per 9 parts water. Keep the
product wet with the bleach solution for 10 minutes, then rinse
thoroughly. Or machine wash in hot water (60°C). Tumble dry
low.
Cushion Care Cautions:
- Valve(s) must be closed. DO NOT allow water or cleaning solution to enter the cushion.
- DO NOT machine wash or sanitize the cushion or machine dry the cushion.
- Use of the following may cause damage to the cushion: abrasives (steel wool, scouring pads); caustic,
automatic dishwasher detergents; cleaning products that contain petroleum or organic solvents, including
acetone, toluene, Methyl Ethyl Ketone (MEK), naphtha, dry cleaning fluid, adhesive remover; oil-based
lotions, lanolin, or phenolic-based disinfectants; ozone gas; ultraviolet light cleaning methods.
To Clean the Cushion: Remove the cover, close valve(s) and place in a large sink. Hand wash, using mild
liquid hand soap, hand dishwashing liquid, laundry detergent, or a multipurpose detergent (Follow product label
instructions.) Gently scrub all cushion surfaces, using a soft plastic bristle brush, a sponge or a cloth. Rinse with
clean water. Air dry. DO NOT expose to direct sunlight.
To Disinfect the Cushion: Repeat cleaning instructions, using 1 part household liquid bleach per 9 parts water.
Keep the product wet with the bleach solution for 10 minutes. Rinse thoroughly with clean water. Air dry.
Note About Sterilization: High temperatures accelerate aging and will damage the product assembly.
ROHO products are not packaged in a sterile condition, nor are they intended or required to be sterilized prior to
use. If institution protocol demands sterilization: First, follow the cleaning and disinfecting instructions; then,
open the inflation valve and use the lowest possible sterilizing temperature, but not to exceed 60°C (158°F),
for the shortest time possible. DO NOT autoclave.