Mechanical cleaning and disinfection
EQUIPMENT
– Cleaning and disinfection device in conformance with
DIN EN ISO 15883
– Cleaning agent
– Neutralising agent if necessary
PROCEDURE
• Arrange the components so that they can be cleaned most
effectively.
• Select a suitable cleaning programme and agent.
VALIDATION
The method has been validated using:
– G7836 CD cleaner-disinfector produced by Miele,
programme Vario TD.
– Standard baskets manufactured by Miele
– Neodisher Mediclean forte 0.5% and Neodisher Z for neutral-
isation (from Dr. Weigert)
DRYING
Even if the cleaner-disinfector is equipped with a drying func-
tion, make sure that no residual moisture is left in the compo-
nents. If necessary, remove any residual moisture by shaking
the parts and allow them to dry completely.
Manual cleaning
As an alternative to mechanical cleaning and disinfection, the
following manual method can be used:
EQUIPMENT
– Cleaning agent
– Brush
– Warm water
- 36 -