Infinite setting
The mechanical zero of galvanometer must be periodically checked. In order perform this checking, be sure that the insulation
tester is off. The pointer should stay on the right end of the scale just over the infinite mark on scale C. In other case, the
plastic screw at the bottom of the galvanometer acrylic cover shall be adjusted.
"
Replacement fuse
Para substituir el fusible debe utilizarse un destornillador de pequeño porte, y girar la tapa del portafusible ) aproximada-
mente un cuarto de vuelta en sentido antihorario, hasta sentir que está destrabado y que el resorte interno empuja hacia
afuera al fusible. Luego de verificar si el mismo está abierto puede substituirse por otro de idénticas características.
Fuse Schurter, model SPT 5 X 20 (Time-lag) 2A/ 250V. High breaking capacity H.
Cleaning
Cleaning of this instrument should be carried out using a soft cleaning anti-static liquid, after verifying that it doesn't affect the
plastic parts used in the case and in the Control Panel of this equipment.
Technical specifications
Test voltages (T.V.)
Insulation test up to
Test
Voltage
A
5 kV
0 - 200
10 kV
0 - 400
15 kV
0 - 400
20 kV
0 - 800
Short-circuit current
Test voltages accuracy
Insulation tester accuracy
Analogue indicator
Safety class
Environmental protection
: 5 kV – 10 kV – 15 kV – 20 kV
: 4.000.000 MΩ
MEASURING INTERVALS (MΩ)
B
100 - 3.000
1.000 - 100.000
200 - 6.000
2.000 - 200.000
300 - 9.000
3.000 - 300.000
400 - 12.000
4.000 - 400.000
: 500 µA
: ±2% of nominal test voltages @ R ≥ 100 GΩ
: Class 2 (±2% of full scale deflection)
: Up to 98 mm scale length, taut band, with mirror (thus avoiding parallax errors)
: Meets the requirements of IEC 61010-1/1990, IEC 61010-1/1992 amendment 2
: IP-54 (with closed lid)
C
C x 10
10.000 - 1.000.000
20.000 - 2.000.000
30.000 - 3.000.000
40.000 - 4.000.000
22
Scale
Output
Multiplier
Resistance
x1
10 MΩ
x2
20 MΩ
x3
30 MΩ
40 MΩ
x4