Sunrise Medical DeVilbiss PulseDose LT Manual Del Usuario página 6

Aparato de regulación compacto
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 12
10
6
5
7
8b
UK Bull Nose
I m p o r t a n t P a r t s
1
8a
4
2 3
8c
German DIN
I m p o r t a n t P a r t s
1. Knob– This knob is used to attach the conserving device to the cylinder.
(Model PD1000 only; not needed on Models PD1000U, PD1000R, or PD1000G)
2. •PulseDose Indicator– Either green or red light illuminates each time the
unit pulses oxygen.
•Normal Battery Indicator– A flashing green light indicates that there is
sufficient battery power.
3. •Low Battery Indicator– A flashing red light indicates that there are 4 - 8
hours of battery life remaining. The low battery indication time may be
reduced when using NiMH batteries.
•Change Battery Indicator– A constant red light indicates that the battery
should be changed immediately. Unit can be used only in continuous flow
mode until new batteries are installed.
4. Cannula Fitting– Use this fitting to attach the cannula to your PulseDose
conserving device.
5. Rotary Selector– When this rotary switch is set to "OFF," the unit is not
using battery power and will not pulse. When this switch is set to one of the
numbers, the unit is on and awaiting inspiration through the nasal cannula
at which time it will dose on every breath. The volume of the oxygen
delivered varies according to which prescription flow setting is chosen. The
final setting on the rotary switch is "CF"; this is the continuous flow
position. In this position oxygen will flow from the cannula fitting at the
default continuous flow rate.
6. Oxygen Contents Gauge–Indicates the remaining pressure in the oxygen
cylinder. When this gauge falls into the red section, you should switch to a
new cylinder.
7. Battery Door– Use only a standard "AA" Alkaline or NiMH batteries.
8. Cylinder Connection– 8a = CGA 870 Pin Index
8b = UK Bull Nose Threaded Cylinder Connection
8c = German DIN Threaded Cylinder Connection
11
.

Publicidad

Tabla de contenido
loading

Productos relacionados para Sunrise Medical DeVilbiss PulseDose LT

Este manual también es adecuado para:

Pd1000Pd1000uPd1000g

Tabla de contenido