Problem
Temperature will not increase
to 106 °C.
Temperature controller display
will not come on.
Section 10 Replacement parts and accessories
Personal injury hazard. Use of non-approved parts may cause personal injury, damage
to the instrument or equipment malfunction. The replacement parts in this section are
approved by the manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Replacement parts
Description
Cable, Polymetron 8315 conductivity sensor
Fuse, temperature controller, 0.5 A (FU-1)
Fuse, heater 120 V, 15 A (FU-2)
Fuse, heater 240 V, 8 A (FU-2)
Heater, 120 V
Heater, 240 V
Power cord kit, SC200 Controller (or Polymetron 9500 controller), North
America
Resin column
Resin column refill bag
Resin beads, 5 gallons
Temperature controller
Possible cause
The boiling point of water is
dependant on the atmospheric
pressure.
The fuse is blown.
W A R N I N G
Solution
If there is a small flow of
condensed water from the
condensate drain, the unit is
operating correctly.
Note: The heater increases the
temperature of the sample to the boiling
point of water, based on local air pressure
and temperature.
Examine the continuity of the
fuses. Refer to
Replace a fuse
on page 25 if blown.
Item no.
08319=A=0005
9525.99.7071
9525.99.7070
9525.99.7072
9525.99.7030
9525.99.7031
9202900
8617600
8617700
8617800
9525.99.7050
English 29