User's guide
Step 1
I
Empty the loading cavity
Before loading a new cartridge, release
any unused gas from the used cartridge
by pressing the button or unscrewing the
applicator cap (DermapenCryo™). Discard
the used cartridge. Please ensure that
the used filter does not remain inside the
cavity.
Step 2
I
Load a new cartridge
The cartridges are packed in peel-pack
blisters with an assembled filter and
cartridge opener included. The filters are
used for protection of the applicator. The
opener punctures the cartridge.
DermapenCryo™ can be used with 8g
cartridges only.
* Warning: release any unused gas from
the used cartridge by pressing the
button or unscrewing the applicator cap.
6
Ensure that the loading cavity is empty
of all residue. Open the peel-pack. Take
out the filter/opener with the tweezers.
Hold
pointed downward. Place the filter in
the instrument and make sure that it is
seated flat at the bottom of the loading
Empty loading cavity
cavity. The trocar of the opener should be
pointed upward. Take the cartridge and
put it into the cavity with the narrow side
down, on top of the filter/opener device.
Step 3
Gently
backend onto the front end. Go to the point
where you feel that the contact between
the assembled filter/opener device and
Blistered cartridge and
assembled filter/opener device
the cartridge is established. Now, position
your hands so that the last turn will be in
an uninterrupted movement. Now, rotate
quickly until you feel that the end of the
rotation is achieved. You may eventually
hear a short hissing sound. The shorter the
hiss, the better!
the
DermapenCryo™'s
tip
I
Puncture the unused cartridge
screw
the
DermapenCryo™'s
NO
end
YES
Installation of the assembled
filter/opener device