Commissioning:
➢
Adapt the suction hose
➢ Connect the mains plug.
➢ Switch on the device and start up
4.2 Automatic mode
➢ Insert the mains plug of the
HEBU oscillating saw at the
vacuum cleaner (1).
➢ Switch on the vacuum cleaner (2).
➢
The lamp lights up and the vacuum cleaner
runs for around 0.5 seconds
as a function check. The vacuum cleaner
then goes to the standby mode (3)
Note:
♦
The vacuum cleaners can be used together with a oscillating HEBU cast saw as well as the autopsy
model.
♦
The vacuum cleaner is automatically switched on and off in the standby mode by the HEBU
oscillating saw, provided you have performed the following points beforehand:
-
Every time the vacuum cleaner is first switched on, it goes to the standby mode.
-
or after being switched on again at least 6 secs. after last being switched off.
♦ The vacuum cleaner has a run-on time of up to 6 secs. This is intended to ensure that the vacuum
cleaner hose is completely emptied.
♦ If the vacuum cleaner is switched on again less than 4 seconds after being switched off,
the suction process operates without automatic function. However the HEBU oscillating saw
can still be switched on and off.
♦ The voltage specification of the vacuum cleaner must agree with the voltage specification
of the HEBU oscillating saw.
3
EN
2
1