Cleaning of vials & sampling containers
Vials, caps and sampling containers should be cleaned thoroughly after each test to avoid
interferences. Minor residuals can cause errors.
Residuals:
The vial must be cleaned according to the type of sample measured.
Replace scratched vials immediately.
Rinse vials thoroughly with deionised water after each measurement.
Clean all glassware thoroughly with laboratory detergent and rinse with deionised water.
Clean heavy contamination by filling the vials with 1:1 HCl followed by multiple rinses with
distilled or deionised water.
Allow vials to air dry.
Touch vials only at the top to minimise dirt and fingerprints.
Wipe the vials with the delivered cleaning cloth to remove waterdrops and fingerprints.