The NEW MAMILAT breast pump does not need maintenance or lubrication.
It is necessary to check functioning and instrument before every use. Unpack the instrument and
always check integrity of plastic parts and feeding cable, they migth have been damaged during
previous use. Connect cable to electrical network and turn switch on.
Close the aspirator outlet with your finger and with suction regulator in maximum vacuum position
check that the vacuum power.
Rotate the knob from right to left and check the aspiration regulating control.
The vacuum power should go down to the minimum value.
Verify that loud noises are not present, these can indicate wrong functioning.
A protection fuse (F 1.6 A 250V and F 4 A 250 V for voltage 110 V / 60 Hz) not reachable from
exterior protects the instrument.
For fuse replacing please contact CA-MI technical assistance.
Fault type
1. No aspiration
2. No aspiration
3. No aspiration due to
outflowing of milk from
overflow bottle
4. The motor doesn't
work (green ligth on the
power switch on)
Faults 1 - 2 - 3 - 4
If the overfill security system it's activated don't proceed with the liquid aspiration.
The bacteriological filter will be stopped the aspiration.
If both the security system doesn't work, there is the possibility that the liquid comes inside the
device, in this case return the device to CA-MI technical service.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING,
CA-MI DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL SERVICE
MAINTENANCE
Cause
Bottle cap badly screwed
down
Cap seal not in its seat
Filter blocked
Activation of the thermal
protection on the motor
None of the remedies has
achieved the desired
results
PLEASE CONTACT CA-MI TECHNICAL SERVICE.
CHECKING, APPEARS TO BE TAMPERED.
Unscrew the cap, then rescrew it
Unscrew the cap and insert the seal
properly in its seal
Wait for some times to avoid the
deactivation of the thermal protection,
after that the motor start again to work
Contact the seller or CA-MI After-sales
Assistance Service
14
Solution
correctly
Replace filter