80i-110s Instruction Manual
15
MAINTENANCE
Before each use, assure continued safety by inspecting the probe. Look for
cracks or missing portions of the probe housing and output cable insulating cover
and for loose or weakened components. Pay particular attention to the insulation
surrounding the probe jaws. If a probe fails this inspection, tape it shut to prevent
unintended operation. To check probe performance, complete the "Performance
Verification" procedure.
WARNING
THESE SERVICING INSTRUCTIONS ARE FOR USE BY QUALIFIED
PERSONNEL ONLY. TO AVOID ELECTRIC SHOCK, DO NOT PERFORM ANY
SERVICING PROCEDURES UNLESS YOU ARE QUALIFIED TO DO SO.
READ THE INFORMATION TITLED "USING THE CURRENT PROBE SAFELY"
AT THE BEGINNING OF THIS USERS MANUAL BEFORE PROCEEDING.
Repairs or servicing not covered in this Users Manual should be performed only
at a Fluke Service Center. A probe under warranty will be promptly repaired or
replaced (at Fluke's discretion) and returned at no charge.
CLEANING AND STORAGE
Periodically wipe the case with a damp cloth and detergent; do not use abrasives
or solvents. Open the jaws and wipe the magnetic pole pieces with a lightly oiled
cloth. Do not allow rust or corrosion to form on the magnetic core ends. If the
probe is not used for periods of longer than 60 days, the battery should be
removed and stored separately.